Radiation Dose Analysis of Non-Contrast Abdominal CT Examination at 90 KV Tube voltage for Kidney Stone Detection
Authors
Department of Physics, Faculty of Science and Mathematics, Diponegoro University, Semarang, Indonesia (Indonesia)
Article Information
DOI: 10.51583/IJLTEMAS.2026.151000068
Subject Category: Biomedical
Volume/Issue: 15/10 | Page No: 830-833
Publication Timeline
Submitted: 2026-10-27
Accepted: 2026-11-01
Published: 2026-11-07
Abstract
Computed Tomography (CT) is a popular modality for detecting kidney stones due to its high-resolution image production. This study aimed to analyze the radiation dose of non-contrast abdominal CT examinations for kidney stone detection at 90 kV. The method used in this study was quantitative analysis using patient data from non-contrast abdominal CT examinations. The parameters analyzed included Computed Tomography Dose Index Volume (CTDIvol) and Dose Length Product (DLP). Statistical analysis was performed using the Shapiro–Wilk test and the independent-samples t-test. The results showed that using a 90 kV voltage produced a CTDIvol value of 2.27 mGy and a DLP of 102.4 mGy•cm. The results of this study indicate that the application of a 90 kV tube voltage in non-contrast abdominal CT examinations for the detection of kidney stones offers clear advantages, both in terms of radiation dose and image quality. The use of a lower voltage consistently reduces radiation exposure to patients—as evidenced by lower CTDIvol and DLP values—with a statistically significant effect.
Keywords
CT abdomen, tube voltage, radiation dose
This article is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).
Downloads
References
1. Bushberg, J. T., Seibert, J. A., Leidholdt, E. M., & Boone, J. M. (2012). The Essential Physics of Medical Imaging. In Wolters Kluwer (3rd ed.). https://doi.org/10.62906/bs.book.174 [Google Scholar] [Crossref]
2. Buri, S., & Shrivastava, V. (2023). A Brief Review of Kidney Stone Detection and Prediction Techniques. In Proceedings of International Conference on Communication and Computational Technologies (pp. 529–543). https://doi.org/10.1007/978-981-99-3485-0_42 [Google Scholar] [Crossref]
3. Eijsvoogel, N. G., Hendriks, B. M. F., Kok, M., Mees, B. M. E., Mihl, C., Wildberger, J. E., & Das, M. (2021). Automated tube voltage selection and adapted contrast media injection protocols in CT angiography of the thoracoabdominal aorta. Chinese Journal of Academic Radiology, 4(2), 116– 124. [Google Scholar] [Crossref]
4. https://doi.org/10.1007/s42058-021-00065-3 [Google Scholar] [Crossref]
5. Gassenmaier, S., Winkelmann, M. T., Magnus, J. P., Brendlin, A. S., Walter, S. S., Afat, S., Artzner, C., Nikolaou, K., & Bongers, M. N. (2022). Low-Dose CT for Renal Calculi Detection Using Spectral Shaping of High Tube Voltage. RoFo Fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren, 194(9), 1012–1019. https://doi.org/10.1055/a-1752-0472 [Google Scholar] [Crossref]
6. Noveranty, A., Purwaningsing, S., & Fendriani, Y. (2024). Analisis Pengaruh Variasi Faktor Eksposi Pada Ct scan Terhadap Kualitas Citra Dan Dosis Radiasi Pada Pemeriksaan Abdomen. Jurnal Pendidikan Fisika Universitas Jambi, 9(3), 53–59. [Google Scholar] [Crossref]
7. Fiarka, F., Oktamuliani, S., & Nuraeni, N. (2023). Pengaruh Tegangan terhadap Nilai CTDIvol, Dose Length Product dan Dosis Efektif pada Pemeriksaan Computed Tomography (CT) Abdomen: Studi di Rumah Sakit Umum Pusat Dr. M. Djamil. Jurnal Fisika Unand, 12(4), 591–597. [Google Scholar] [Crossref]
8. https://doi.org/10.25077/jfu.12.4.591-597.2023 [Google Scholar] [Crossref]
This work is licensed under CC BY 4.0